2553A Precision DC Calibrator

Brand: Yokogawa Electric Corporation

The 2553A provides popular K, T, E, J, and an additional 6 types of thermocouples.

These 10 types are all regulated in IEC. It also provides widely used Pt100 as an RTD.

Resistance as well as temperature value can be set as an output value, enabling instrument calibration by both values.
Features
  • Digital Display of Output
  • High Basic Accuracy 
  • High Stability and Low Noise 
  • High Resolution 

Specifications

Accuracy

  • ± (40 ppm of Setting + 4 µV) to ± (40 ppm of Setting + 300 µV) for 180 Days - Voltage Generation
  • ± (50 ppm of Setting + 0.03 µA) to ± (70 ppm of Setting + 4 µA) for 180 Days - Current Generation
  • ± (55 ppm of Setting + 0.015 ohms) for 180 Days - Resistance Generation
  • ± (60 ppm of Setting + 4 µV) to ± (60 ppm of Setting + 450 µV) for 1-Year - Voltage Generation
  • ± (75 ppm of Setting + 0.015 ohms) for 1-Year - Resistance Generation
  • ± (80 ppm of Setting + 0.04 µA) to ± (100 ppm of Setting + 5 µA) for 1-Year - Current Generation
  • ± 0.1° C (± 0.18° F) for 180 Days - RTD
  • ± 0.15° C (± 0.27° F) for 1-Year - RTD
  • ± 120 ppm (1-Year) for 1 mA Range - DC Current
  • ± 75 ppm (1-Year) for 1 V Range - DC Voltage

Approvals & Certifications

  • AS
  • BS
  • CSA
  • GB
  • NBR
  • UL
  • VDE

Communication

  • Baud Rate: 100 Mbps
  • Ethernet
  • GPIB
  • USB

Consumption

  • 30 VA - Max.

Current

  • 0.1 μA/V or more (DC, 50/60 Hz) - CMRR
  • 0.5-2 mA - Nominal Current
  • 1-100 mA - DC Current
  • 30 mA - Max.

Display

  • 5-1/2-Digits
  • 7-Segment LED's

Electrical Connectors

  • RJ45 Connector
  • Typical B Connector

Frequency

  • 48 to 63 Hz - Allowable Power Supply Frequency Fluctuation Range
  • 50/60 Hz

Humidity

  • 20 to 80 % Relative Humidity

Load

  • 1 MH - Max.
  • 10 μF - Max.

Mounting

  • Rack-Mounted

Operating Temperature

  • 5° to 40° C (41° to 104° F)

Output

  • 4-20 mA, 2-Wire
  • 80 to 100 mV

Output Noise

  • 0.015-0.3 μArms for 0.1 to 10 Hz - Current Generation
  • 0.5-10 μArms for 10 Hz to 10 kHz - Current Generation
  • 10-50 μVrms for 10 Hz to 10 kHz - Voltage Generation
  • 1-20 μVrms for 0.1 to 10 Hz - Voltage Generation
  • 2 μVrms - Low Noise

Power Supply

  • 100-120 Vac
  • 200-240 Vac

Resistance

  • ≤ 10 megaohms - Output Resistance for 30 mA & 100 mA
  • ≤ 10 milliohms - Output Resistance for (100 mV-30 V Range)
  • ≤ 100 megaohms - Output Resistance for 1 mA & 10 mA
  • 400 ohms Range - Resistance Generation
  • Approximately 1 ohms - Output Resistance for 10 mV Range

Resolution

  • 0.01 ohms - Resistance Generation
  • 0.01-1 μA - Current Generation
  • 0.1° C (32.18° F) - Temperature Generation for RTD
  • 100 NV-1 mV - Voltage Generation
  • 120,000 Count - Output Resolution

Response Time

  • 500 milliseconds

RTD Type

  • Pt100

Sensor, Probe Type

  • A Thermocouple
  • B Thermocouple
  • C Thermocouple
  • E Thermocouple
  • J Thermocouple
  • K Thermocouple
  • N Thermocouple
  • RTD
  • R Thermocouple
  • S Thermocouple
  • T Thermocouple

Stability

  • ± 15 ppm/h - High Stability

Storage Temperature

  • -15° to 60° C (5° to 140° F)

Supply Voltage

  • 1,500 Vac/1 Minute - Withstand Voltage
  • 10 mV-30 V - DC Voltage
  • 180-264 Vac - Allowable Power Supply Voltage Fluctuation Range
  • 90 to 132 Vac - Allowable Power Supply Voltage Fluctuation Range

Temperature Coefficient

  • ± (10 % of Setting + 0.1 μV)/°C - for 10 mV Voltage Generation Range
  • ± (10 % of Setting + 0.15 μA)/°C - for 100 mA Current Generation Range
  • ± (3 % of Setting + 0.0015 μA)/°C - for 1 mA Current Generation Range
  • ± (3 % of Setting + 1.5 μV)/°C - for 1 mV Voltage Generation Range
  • ± (3 % of Setting + 15 μV)/°C - for 10 V Voltage Generation Range
  • ± (3 % of Setting + 45 μV)/°C - for 30 V Voltage Generation Range
  • ± (5 % of Setting + 0.015 μA)/°C - for 10 mA Current Generation Range
  • ± (5 % of Setting + 0.3 μV)/°C - for 100 mV Voltage Generation Range
  • ± (7 % of Setting + 0.045 μA)/°C - for 30 mA Current Generation Range
  • ± 0.002 ohms/°C - Resistance Generation
  • ± 0.006° C/°C - Temperature Generation for RTD

Temperature Range

  • -200° to 850° C (-328° to 1,562° F) - Source Range for RTD (Pt100)
  • -270° to 2,500° C (-454° to 4,532° F) - Source Range for Thermocouple Typicals (R, S, B, J, T, E, K, N, C, A)

Update Time

  • 30 minutes - Warm-Up Time
show more