Accuracy
- ± (40 ppm of Setting + 4 µV) to ± (40 ppm of Setting + 300 µV) for 180 Days - Voltage Generation
- ± (50 ppm of Setting + 0.03 µA) to ± (70 ppm of Setting + 4 µA) for 180 Days - Current Generation
- ± (55 ppm of Setting + 0.015 ohms) for 180 Days - Resistance Generation
- ± (60 ppm of Setting + 4 µV) to ± (60 ppm of Setting + 450 µV) for 1-Year - Voltage Generation
- ± (75 ppm of Setting + 0.015 ohms) for 1-Year - Resistance Generation
- ± (80 ppm of Setting + 0.04 µA) to ± (100 ppm of Setting + 5 µA) for 1-Year - Current Generation
- ± 0.1° C (± 0.18° F) for 180 Days - RTD
- ± 0.15° C (± 0.27° F) for 1-Year - RTD
- ± 120 ppm (1-Year) for 1 mA Range - DC Current
- ± 75 ppm (1-Year) for 1 V Range - DC Voltage
Approvals & Certifications
Communication
- Baud Rate: 100 Mbps
- Ethernet
- GPIB
- USB
Current
- 0.1 μA/V or more (DC, 50/60 Hz) - CMRR
- 0.5-2 mA - Nominal Current
- 1-100 mA - DC Current
- 30 mA - Max.
Display
- 5-1/2-Digits
- 7-Segment LED's
Electrical Connectors
- RJ45 Connector
- Typical B Connector
Frequency
- 48 to 63 Hz - Allowable Power Supply Frequency Fluctuation Range
- 50/60 Hz
Humidity
- 20 to 80 % Relative Humidity
Operating Temperature
- 5° to 40° C (41° to 104° F)
Output
- 4-20 mA, 2-Wire
- 80 to 100 mV
Output Noise
- 0.015-0.3 μArms for 0.1 to 10 Hz - Current Generation
- 0.5-10 μArms for 10 Hz to 10 kHz - Current Generation
- 10-50 μVrms for 10 Hz to 10 kHz - Voltage Generation
- 1-20 μVrms for 0.1 to 10 Hz - Voltage Generation
- 2 μVrms - Low Noise
Resistance
- ≤ 10 megaohms - Output Resistance for 30 mA & 100 mA
- ≤ 10 milliohms - Output Resistance for (100 mV-30 V Range)
- ≤ 100 megaohms - Output Resistance for 1 mA & 10 mA
- 400 ohms Range - Resistance Generation
- Approximately 1 ohms - Output Resistance for 10 mV Range
Resolution
- 0.01 ohms - Resistance Generation
- 0.01-1 μA - Current Generation
- 0.1° C (32.18° F) - Temperature Generation for RTD
- 100 NV-1 mV - Voltage Generation
- 120,000 Count - Output Resolution
Sensor, Probe Type
- A Thermocouple
- B Thermocouple
- C Thermocouple
- E Thermocouple
- J Thermocouple
- K Thermocouple
- N Thermocouple
- RTD
- R Thermocouple
- S Thermocouple
- T Thermocouple
Stability
- ± 15 ppm/h - High Stability
Storage Temperature
- -15° to 60° C (5° to 140° F)
Supply Voltage
- 1,500 Vac/1 Minute - Withstand Voltage
- 10 mV-30 V - DC Voltage
- 180-264 Vac - Allowable Power Supply Voltage Fluctuation Range
- 90 to 132 Vac - Allowable Power Supply Voltage Fluctuation Range
Temperature Coefficient
- ± (10 % of Setting + 0.1 μV)/°C - for 10 mV Voltage Generation Range
- ± (10 % of Setting + 0.15 μA)/°C - for 100 mA Current Generation Range
- ± (3 % of Setting + 0.0015 μA)/°C - for 1 mA Current Generation Range
- ± (3 % of Setting + 1.5 μV)/°C - for 1 mV Voltage Generation Range
- ± (3 % of Setting + 15 μV)/°C - for 10 V Voltage Generation Range
- ± (3 % of Setting + 45 μV)/°C - for 30 V Voltage Generation Range
- ± (5 % of Setting + 0.015 μA)/°C - for 10 mA Current Generation Range
- ± (5 % of Setting + 0.3 μV)/°C - for 100 mV Voltage Generation Range
- ± (7 % of Setting + 0.045 μA)/°C - for 30 mA Current Generation Range
- ± 0.002 ohms/°C - Resistance Generation
- ± 0.006° C/°C - Temperature Generation for RTD
Temperature Range
- -200° to 850° C (-328° to 1,562° F) - Source Range for RTD (Pt100)
- -270° to 2,500° C (-454° to 4,532° F) - Source Range for Thermocouple Typicals (R, S, B, J, T, E, K, N, C, A)
Update Time