SIREC D200 Process Recorder
Brand: Siemens AGThe SIREC D200 is a 144 mm x 144 mm display recorder with 3 to 12 analog channels and up to 8 digital channels.
The wide range of SIREC D products accommodates continuous monitoring, system maintenance, process optimization and troubleshooting requirements.
Specifications
Brand
Display Size and Type
- 5.7” diagonal, color Diagonal, Digital Colour LCD (TFT) with Touch Screen Industrial Grade with brightness adjustment and wide viewing angle
Resolution
- QVGA (320 x 240 pixels)
Screen Saver
- Set in minutes from 1 … 720, can be set to dim the screen or to switch off. Automatic wake-up facility in the case of an alarm
Brightness Adjustment
- Adjustable between 10 and 100 %, default set to 80 % brightness
Backlight Life Time
- 50 000 hours to half brightness when used at 100 %. Max. luminosity 450 cd/m²
Touch Screen Life
- 1 000 000 touches
Display Update Rate
- Display values updated every second
Status Display
- A status bar, at the top of the recorder’s screen, displays the realtime icons of the recorder status, such as recording time left and alarm active.
Communications
- Ethernet 10/100 base - T with RJ45 connector supporting Modbus/TCP, FTP, Internet, DHCP or fixed IP address.
Mathematics
- Basic maths include Add, Subtract, Multiply, Divide, Modulo and power. Full Maths (option) support up to 100 character free form math expression for each pen. Like SINE, COS, TAN, Log, Parenthesis (eg. A1 + A2), comm variables, free memory, and access to any data item variable (A1, P1, D1 etc.)
Front USB Port
- Front USB host port for data and setup transfers. External devices keyboard or mouse, Barcode reader, or external mass storage device. (USB 2.0 compliant)
Standard Screens
- Digital values displayed include: • Alarms on bars • Engineering units • Pen name • Tag, time and date • 20 character description • Totalised values
Local Mass Storage Options
- USB memory key - must be formatted
- USB hard drive - up to 120 GByte
Internal Data Buffer
- Non-volatile, 1 GByte (9 230 million acquisition values) and 2 GByte (up to 650 million points)
Setup and Screens
- Stored internally on non-volatile memory
Manual Saving
- Data saving by inserting USB memory stick
Data Saving Period
- Related to log rate, number of pens, totals and alarms. Each pen is capable of its own independent storage rate (100 ms … 60 h)
Data Format
- Binary encoded format
Recycling Mode
- Internal memory has FIFO (First In First Out) capability where the newest data over-writes the oldest data
Voltage (VRMS)
- 00 Vac … 250 Vac (Auto select)
Frequency
- 50/60 Hz
Power Consumption
- < 40 W
24 V Optional Instrument Power
- 12 ... 30 Vdc / 12 ... 20 Vac Power Consumption: < 40 W
Battery
- Battery backed up for clock, replaceable lithium battery Type 6032, 3.0 V - 10 years life (Recorder powered), 4 years life, typical (Recorder unpowered).
Password Protection
- Multiple Administrator control of password setup and management with four levels of password protection for - Engineer, Supervisor, Technician, and Operator. Up to 50 different users are available. Password protection restricts user entry to the recorder set up and specific screens.
Engineer
- Highest access to all levels, Supervisor, Technician and Operator
Supervisor
- 2nd highest level including Technician and Operator access
Technician
- 3rd level including Operator access
Operator
- 4th and lowest level of access
Temperature Units
- °C, °F, K
Recorder Identification
- Status bar: Alternately Displays Recorder ID and Recorder Screen Name, Displays Time and Date
Clock Accuracy
- ± 20 ppm (± 1 minute/month) at 25° C
Alarm Set Points
- 6 per pen integral ”soft” alarm set points easily set by user to announce selected out of limit conditions, user can select if an alarm triggers a change in the chart background colour
Alarm Triggers
- Alarm triggers can be set for Hi, Lo, Deviation. Latched alarms require acknowledgement from the operator
Alarm Damping
- 1 s … 24 h
Hysteresis
- ± 100 % of pen scale An alarm can change the log rate on the affected pen
Data Replay Mode
- Data replay facility on chart displays at normal, fast or slow speeds with zoom and cursor. Jump facility from the alarm history list directly to the occurance on the chart
Chart Rates
- • 1 mm/h • 5 mm/h • 10 mm/h • 20 mm/h • 30 mm/h • 60 mm/h • 120 mm/h • 600 mm/h • 1200 mm/h • 6000 mm/h
Messages Screen
- The message screen displays system information and records any setup activity that has been changed. It also provides warning and error message updates, lists alarm activity and will display user defined marks on a chart
CE Conformity (CE Mark)
- This product conforms with the protection requirements of the following European Council Directives: 2014/35/EC, the Low Voltage Directive, and 2014/30/EC, the EMC Directive. Conformity of this product with any other ”CE Mark” Directive (s) shall not be assumed.
Immunity Product Classification
- Complies with EN 61326-1: 2013 Class I: Cord Connected, Panel Mounted Industrial Control Equipment with protective earthing (grounding), (EN 61010-1: 2010)
Enclosure Rating
- Front panel designed to NEMA 3/IP54 (Optional NEMA 4X/IP66)
Installation Requirements
- Category II: Overvoltage (EN 61010-1: 2010) Pollution Degree 2
EMC Standards
- Emissions - EN 61326-1: 2013 Class A Immunity - EN 61326-1: 2013 Industrial Levels
Safety
- Complies with EN 61010-1: 2010 Panel Mounted Equipment, Terminals must be enclosed within the panel
Number of Inputs
- 3, 6, 9 or 12 input channels
Input Types
- mV, V, mA with external shunt (provided as standard), Thermocouple, RTD and ohms
Minimum Input Span
- Range is fully configurable with span limitation of the operating range selected with 4 % under range to 4 % over-Range capability (50 V Range 2 %)
Burnout (T/C)
- Active (High or Low), Passive and Health watch/maintenance (option)
Cold Junction Compensation
- Internal compensation with the ability to manually adjust values, External Input for compensation, External CJC value specified
Input Resolution
- 0.0015 % (16 Bit ADC)
Current Loop Resistance
- 10 Ω , Use ± 0.1 % External Resistor, Volts > 1 MΩ, All Other > 10 MΩ
T/C and RTD
- 100 Ω per lead Max. (a single point cal on Slot A will improve accuracy for a lead resistance above 10 Ω)
Square Root Extraction
- Available as standard on Volts and mA input types
Sensor Compensation
- Single point and Dual point for every input type
Input Sampling Rate
- Recorder has 2 available slots with up to 6 analog inputs each, first slot fixed
Analog Input Card (Standard)
- 200 ms (5 Hz), 500 ms (2 Hz)
Analog Input Expansion Card (Option)
- 100 ms (10 Hz), 200 ms (5 Hz), 500 ms (2 Hz)
Linear Scales
- Normal and Scientific Notation
- Decimal Point automatic or programmable
- Engineering units, user definable (10 characters)
Logarithmic Decade Limits
- -38 min … +38 max, (recommend up to 20 decades on one screen to ensure clarity)
Input Isolation
- 300 Vac channel-to-channel, channel-to-ground (Resistance thermometers are not isolated for initial card, expansion card option RTs are isolated)
Noise Rejection
- At 50/60 Hz ± 2 %
Input Actuation (Linear)
- Range
mV (DC)
- -1000 … +1000
V (DC)
- -50 … +50
mA
- 4 … 20, 0 … 20
200 Ω
- 0 … 200
500 Ω
- 0 … 500
1000 Ω
- 0 … 1000
4000 Ω
- 0 … 4000
Thermocouples
- Temperature range
Thermocouples B
- 260 ... 538° C (500 ... 1000° F)
- 538 ... 1816° C (1000 ... 3300° F)
Thermocouples E
- -270 ... -200° C (-454 ... -328° F)
- -200 ... -70° C (-328 ... -94° F)
- -70° to +1000° C (-93° to +1832° F)
Thermocouples J
- -210° ... 0° C (-346° ... +32° F)
- 0° ... 1200° C (32° ... 2192° F)
Thermocouples K
- -270 ... -70° C (-454 ... -94° F)
- -70° to +1372° C (-93° to +2502° F)
Thermocouples R
- -50° to +260° C (-58° to +500° F)
- 260 ... 650° C (500 ... 1202° F)
- 650 ... 1768° C (1202 ... 3214° F)
Thermocouples S
- -50° to +260° C (-58° to +500° F)
- 260 ... 1000° C (500 ... 1832° F)
- 1000 ... 1768° C (1832 ... 3214° F)
Thermocouples T
- -270 ... -210° C (-454 ... -346° F)
- -210° ... +400° C (-346° ... +752° F)
Thermocouples L
- -200 ... 0° C (-328° to +32° F)
- 0° ... 900° C (32° ... 1652° F)
Thermocouples G (W_W26)
- 0° ... 100° C (32° ... 212° F)
- 100 ... 316° C (212 ... 600° F)
- 316 ... 830° C (600 ... 1526° F)
- 830 ... 1515° C (1526 ... 2759° F)
- 1515 ... 2315° C (2759 ... 4119° F)
Thermocouples C (W5, W26)
- 0° ... 180° C (32° ... 356° F)
- 180 ... 1220° C (356 ... 2228° F)
- 1220 ... 2315° C (2228 ... 4199° F)
Thermocouples M (NiMo-NiCo) (NNM90)
- -50° to +370° C (-58° to +698° F)
- 370 ... 1410° C (698 ... 2570° F)
Thermocouples N (Nicosil Nisil)
- -200° to +100° C (-328° to +212° F)
- 100 ... 1300° C (212 ... 2372° F)
Thermocouples Chromel/Copel
- -50° to +600° C (-58° to +1112° F)
Thermocouples P (Platinel)
- 0° ... 1390° C (32° ... 2534° F)
Thermocouples D
- 0° ... 180° C (32° ... 356° F)
- 180 ... 1840° C (356 ... 3344° F)
- 1840 ... 2490° C (3344 ... 4515° F)
Pt100 α = 0,00385
- -200° to +850° C (-328° to +1562° F)
Pt 200 α = 0.00385
- -200° to +850° C (-328° to +1562° F)
Pt 500 α = 0.00385
- -200° to +850° C (-328° to +1562° F)
Pt 1000 α = 0.00385
- -200° to +850° C (-328° to +1562° F)
Nickel, 100 Ω
- -60° ... +180° C (-76° ... +356° F)
Nickel, 120 Ω
- 80° to +260° C (-112° to +500° F)
Logging Method
- Sample, Average, Min/Max. - can be set independently per pen
Logging Types
- Continuous, Fuzzy
Logging Rate
- From 100 ms … 60 h per Pen
Fuzzy Logging
- A secure data storage technique which delivers data compression ratio of 100: 1 or more, self teaching, storing the data at a variable rate to match the process
Enclosure/Bezel
- Zinc-Plated Steel case with high impact resistant polycarbonate bezel, scratch resistant lens (Polyethylene Terephthalate). NEMA 3/IP54 protection rating standard. Optional NEMA 4X/IP66 (Front face only)
Colour
- Bezel: Grey
Mounting Pane
- Unlimited mounting angle For the best view of the display the viewing angle should not exceed: • 45° from the left or right, • 10° looking down and • 30° looking up at the recorder display.
- Mounting adjustable for panel thickness of 2 mm … 20 mm. Adapter kits available
Dimensions (W x H x D) in mm
- 144 x 144 x 200 (5.67 x 5.67 x 7.87”)
- Additional 80 mm (3.15”) clearance recommended for a straight type power cable and signal connectors
Cutout (W × H) in mm
- 138 x 138 mm (5.43 x 5.43”)
Weight
- Maximum 2.4 kg (5.3 lb)
Wiring Connections
- IEC Power Plug. Removable terminal strip for input and alarm connections
Ambient Temperature
- 0° C … 50° C (32° F to 122° F)
Relative Humidity (%RH)
- 10 ... 90
Vibration Frequency (Hz)
- 0 ... 70
Vibration Acceleration (g)
- 0.1
Mechanical Shock Acceleration (g)
- 1
Mechanical Shock Duration (ms)
- 30
Mounting Position from Vertical Tilted Forward
- 40°
Mounting Position from Vertical Tilted Backward
- 65°
Mounting Position from Vertical Tilted to Side (±)
- 65°
Power Requirements Mains Voltage (Vrms)
- 100 … 250
Power Requirements Low Voltage AC (Vrms)
- 20 … 25
Power Requirements DC Voltages
- 20 … 30
Power Requirements Frequency (Hz)
- 47 … 63
Warm Up
- 30 minutes minimum